Description
Adash America A4300 RU Run up Module
Measurement Control When you need to measure Run Up or Coast Down of the machines, then you should use the Runup mode. It enables the same measurements as in Analyzer mode, but controlled by speed, time or uncontrolled. What does it mean “controlled”? In Analyzer mode you can measure the Set and then you have to save measurements manually.
The Runup is different. The measurements are saved automatically and additionally the measurements are regularly repeated under the control. The trigger function is used for that control. The menu item Menu/Trigger/Runup mode is used for setting.
Usually the speed is used for that control and new measurement is made, when the speed changes from previous measurement of defined value (e.g.10 RPM).
Also the time can be used to control that procedure. Then you can take the measurement in defined time interval ( e.g. each 60 sec ).
The asap is the next choice. It means no delay between measurements – As Soon As Possible. Be careful for such mode. You can use all memory quickly.
The last choice is the Manual mode. You can press the Ok button at any suitable moment to make a new measurement.
Available interval is displayed on the status bar before the start of measuring. The instrument enables save up to 1GB of data on each measurement (if enough memory available), but the limit of 1 million values per set cannot be exceeded.