Description
Keysight 34907A – Multifunction Module for 34970A/34972A
The Keysight 34907A allows great flexibility for a variety of sense and control applications. It combines two 8-bit ports of digital input and output, a 100 kHz gated totalizer, and two ± 12 V analog outputs—all on a single earth-referenced module. The digital inputs and totalizer input may be included in a scan. Alarm limits for the digital and event counter inputs are evaluated continuously, capturing and logging alarm conditions even between scans.
Digital input/output
Use the digital outputs with an external power supply to control microwave switches and attenuators, solenoids, power relays, indicators, and more. Use the digital inputs to sense limit switch and digital bus status. There are no complex handshake modes; reads and writes are initiated either from the front panel or the bus.
- Port 1, 2: 8 bit, input or output, nonisolated
- Vin(L): < 0.8 V (TTL)
- Vin(H): > 2.0 V (TTL)
- Vout(L): < 0.8 V @ Iout = -400 mA
- Vout(H): > 2.4 V @ Iout = 1 mA
- Vin(H) max: < 42 V with external open drain pull-up
- Alarming: Maskable pattern match or state change
- Speed: 4 ms (max) alarm sampling
- Latency: 5 ms (typical) to 34970A alarm output
- Read/write speed: 95/s
Totalize input
Count events from devices like photo interrupters, limit switches, and Hall-effect sensors.
It keeps an updated total which can be read via the front panel or programmatically at any time. With 26 bits of resolution, it can count events at full speed for nearly 11 minutes without an overflow.
- Max count: 2²⁶ – 1
- Totalize input: 100 kHz (max) Rising or falling edge, programmable
- Signal level: 1 Vp-p (min) 42 Vpk (max)
- Threshold: 0 V or TTL, jumper selectable
- Gate Input: TTL-Hi, TTL-Lo, or none
- Count reset: Manual or Read + Reset
- Read speed: 85/s
Analog output
Use the two electronically calibrated analog outputs to source bias voltages to your device under test, to control your analog programmable power supplies, or use the outputs as setpoints for your control systems. The outputs are programmed directly in volts, either from the front panel or from the bus.
- DAC 1, 2: ± 12 V, non-isolated
- Resolution: 1 mV
- IOUT: 10 mA max
- Settling time: 1 ms to 0.01% of output
- Accuracy ± (% of output + mV): 1 year ± 5 °C 0.25% + 20 mV
- Temp. coefficient: ± (0.015% + 1 mV)/°C


Reviews