Hipot Tester (AC/DC/IR) – direct replacement for Quadtech Sentry 30 Plus

  • Nist Calibration included
  • 19073 Hipot Tester (AC/DC/IR) – Sentry 30 Plus
  • AC/DC/IR 3 in 1 hipot tester
  • AC 5kV and DC 6kV output
  • 1kV insulation resistance test
  • Insulation resistance measurement from 1M to 50G
  • Ground continuity check
  • Open Short Check(OSC) function
  • GFI shutdown the instrument when imbalance current > 0.5mA
  • Flashover (ARC) detection
SKU: Chroma 19073 Tags: , Categories:
Pay safely with Visa Pay safely with Master Card Pay safely with PayPal Pay safely with Maestro Tooltip text
Guarantee Safe and Secure Payment Checkout

Description

Chroma 19073 Hipot Tester (AC/DC/IR) – Sentry 30 Plus

  • Nist Calibration included
  • 19073 Hipot Tester (AC/DC/IR) – Sentry 30 Plus
  • AC/DC/IR 3 in 1 hipot tester
  • AC 5kV and DC 6kV output
  • 1kV insulation resistance test
  • Insulation resistance measurement from 1M to 50G

Complete Dielectric Testing Solution

The 19050 series electrical safety testers are advanced digital hipot testers with load and line regulation to ensure measurement integrity. Multi-step capability allows users to perform multiple tests in sequence, such as AC hipot followed by IR.

The Chroma Hipot Tester 19050 series provides 3 models to choose from. The 19052 includes AC/ DC/IR Hipot testing and insulation resistance (IR) measurements. The 19053 provides 8 scan channels for IR measurement, and the 19054 provides 4 scan channels for IR measurement in a single compact unit.

The Chroma Hipot Tester 19070 series provides 2 models to choose from. The 19071 is for AC Hipot testing. The 19073 combines both AC and DC Hipot with insulation resistance (IR) measurements into a single compact unit.

Open Short Check (OSC)

The O.S.C function is used to check whether the connection is open, or a short circuit, between the instrument and DUT (device under test) before the Electrical Safety Test, is performed. If the connection is bad between the instrument and DUT, like bad leads or relay oxidation, the test may result in a PASS determination. In some cases, the DUT has a short before testing. Testing continually leads to instrumentation damage due to it suffering the high load current. Therefore, we check for the open and short circuit to ensure the test is effective and to protect the instruments.

Generally, DUT has a capacitive load (Cx) from tens to thousands of pF. If the connection is open, a capacitance will appear and then the total capacitive load is lower than that of normal conditions. If the DUT is shorting, the total capacitive load is higher than that of normal conditions. Therefore, we can measure the value of the capacitive load to verify whether the contact is good or not.

Ground Continuity Check

All of the 19050 series testers have a ground continuity check feature to determine the resistance between the ground blade of the power cord and any exposed metal on the product. This resistance should be less than 1Ω.

Ground Fault Interrupt (GFI)

A requirement for test environments indicates that test equipment is equipped with an auto interrupt device. Chroma builds in a Ground Fault Interrupt (GFI) function for this purpose. When the current meter A1 and A2 detect a difference (i2-i1=iH) between the value i1 and actual i2 test current over high, this device can cut the power transiently for the protection of the operator. It is not only compliant with the safety standard but also provides additional safeguards for test personnel.

Panel Description

1. LINE Switch

2. Window Display

3. Stop Button

4. Start Button

5. Function Keys (F1~F4)

6. Pass Indicator

7. Fail Indicator

8. Test Indicator

9. HV Output

10. RTN/LOW

11. 8 channels HV Output (19053 only)

12. Remote I/O

13. LINE Voltage Selector

14. Power Cord Receptacle

15. RTN/LOW

16. GPIB/Printer Interface (Option)

17. Scan Interface (Option)

18. Fan

19. Remote Interface

20. RS-232 Interface

21. Continuity Test O/P

Safety Testing / Surge Testing/Hipot Tester Template

Test Type HipotAC, DC, IR
Max AC Test Voltage5 kV
Max DC Test Voltage6 kV
Min AC Test Voltage0.05 V
Min DC Test Voltage0.05 V
Max Trip Current AC Test20 mA
Min Trip Current AC Test0.1 mA
Max Trip Current DC Test5 mA
Min Trip Current DC Test0.01 mA
Programmable (hipot)Yes
Ground Continuity Current AC0.5 mA (0.0005 A) 
Insulation Resistance TestYes
IR Test Voltage1,000 V 
Connection TypeProbe
Arc DetectionYes
Maximum Insulation Resistance50 G Ohms (50000 M Ohms)

Product General Attributes

Warranty2 YEARS
Warranty Details2-Year factory warranty
Safety ApprovalCE, UL 
Interfaces I/O RS-485
StorageInternal Memory 
Product Weight26.46 LBS
Product Height4.13 IN
Product Length13.78 IN
Product Width10.63 IN
Shipping Weight30 LBS
Data LoggingYes
HTS/Schedule B Number9030.33.3400
ECCN Number3A992
Calibration IncludedFactory Calibration, NIST Calibration
Country of OriginTaiwan

Reviews

There are no reviews yet.

Be the first to review “Chroma 19073 Hipot Tester (AC/DC/IR) – Sentry 30 Plus”
Weight26.46 lbs