Description

Compliance MegaPulse Defib-5PF2
- MegaPulse Defib-5PF-002 Defibrillation-proof tester with 10 channels multiplexer outputs
- Output voltage: 0-5000 V tolerance +/- 1%
- Main capacitance: 0.13 uF +/-10% Type A, 0.50 uF +/-10% Type B (600,000 cycles life)
- Main resistance: 100Ω, 50Ω, and 400Ω +/- 1% non-inductive
- Duty Cycle: 1 pulse every 20sÂ
Defib-proof and Energy Reduction Tester with 10 Channel Mux.
The MegaPulse tester provides impulses up to 4.46kV peak 1.7kΩ load. The tester has positive and negative waveform delivery via a front panel button. Tester impedance is 39ohms. The MegaPulse tester has voltage waveform 500µs rise time, 3.4ms duration and up to 4.46kV on a 1.7kΩ load.
The MegaPulse impulse tester generates the impulse waveform only; it does not determine Passing or Failing results. It is the operator’s responsibility to monitor the output waveform and determine Passing or Failing results. In monitoring the impulse waveform, consider the following points: The Impulse waveform is high voltage and high frequency (short duration). Always ensure that the measuring instrument (usually an oscilloscope with a high-voltage probe) is rated for the voltage involved, and that the frequency response of the instrument and probe are capable of measuring the output waveform of the MegaPulse Impulse Tester. A measuring instrument or probe with a low frequency response will result in erroneous readings that could be mis-read.
Determination of Passing and Failing results can prove difficult. To obtain the most accurate results, it is generally necessary to perform multiple impulse tests on a few different test samples (that have adequate insulation to pass the impulse test). Take note of the impulse waveshape, amplitude, and duration. Also note how much variance there is in the waveshape from test to test. Also (if possible), perform impulse testing on some test samples that are known to have inadequate (or damaged) insulation. Take note of the impulse waveshape, amplitude, and duration, when an insulation breakdown occurs.
MegaPulse Defib-5PF-002
- Output Voltage: 0 – 5000 V tolerance +/-1%
- Main Capacitance: 0.13 uF +/-10% Type A, 0.50 uF +/-10% Type B (600,000 cycles life)
- Voltage Control: Digital Set point adjusted by frontal VOLTAGE knob or by PC TestMinder
- Voltage Display: 4 Digit LED Display
- Voltage Meter resolution: 2V
- Duty Cycle: 1 pulse every 20s
- Selectable Input Voltage: 120V – 60Hz; 240V – 50Hz

Front Panel:
1. Voltage adjust knob – adjust. set point of tester
2. Polarity switch
3. NOR REV indicator – state of polarity switch
4. Charge switch – starts the charge process
5. Charge indicator
6. Trigger switch – triggers output impulse waveform
7. Trigger indicator
8. Voltage meter – displays output voltage set point
9. 500:1 BNC – 500:1 reference
10-19. Outputs – 10 channels multiplexer outputs

Rear Panel:
1. Input switch
2. Fuse B
3. Appliance inlet
4. Fuse replacement warning
5. Interlock switch – Emergency stop
6. RS-232 Interface (optional)
7. FansÂ
Product General Attributes | |
---|---|
Unique Features | Increased accuracy in accordance with IEC 60601-2-25 and -2-27:2011 requirements |
Product Weight | 50 LBS |
Product Height | 11 IN |
Product Length | 17 IN |
Product Width | 17 IN |