Compliance MegaPulse Defib-5PF4 – Defib-proof and Energy Reduction Tester, Type A Type B Outputs

10 people are viewing this product right now

Shipping calculated at checkout.
Add to quote
Estimated delivery:3 days

Defib-proof and Energy Reduction Tester

  • MegaPulse Defib-5PF-004 Defibrillation-proof tester with ( 4) Type A and (6)Type B outputs
  • Output voltage: 0-5000 V tolerance +/- 1%
  • Main capacitance: 0.13 uF +/-10% Type A, 0.50 uF +/-10% Type B (600,000 cycles life)
  • Main resistance: 100Ω, 50Ω, and 400Ω +/- 1% non-inductive
  • Duty Cycle: 1 pulse every 20s
  • Voltage waveform 500µs rise time, 3.4ms duration and up to 4.46kV on a 1.7kΩ load
  • External interlock, disables HV output and defeats test when circuit is open
  • Direct Joules readout on front panel after Energy Reduction Test
  • Increased accuracy in accordance with IEC 60601-2-25 and -2-27:2011 requirements
  • Dims: 17in wide x 16in high x 17in deep; 50lbs approx.
SKU: Compliance West MegaPulse Defib-5PF4 Tags: , Categories:
Pay safely with Visa Pay safely with Master Card Pay safely with PayPal Pay safely with Maestro Tooltip text
Guarantee Safe and Secure Payment Checkout

Description

Compliance MegaPulse Defib-5PF4

  • MegaPulse Defib-5PF-004 Defibrillation-proof tester with ( 4) Type A and (6)Type B outputs
  • Output voltage: 0-5000 V tolerance +/- 1%
  • Main capacitance: 0.13 uF +/-10% Type A, 0.50 uF +/-10% Type B (600,000 cycles life)
  • Main resistance: 100Ω, 50Ω, and 400Ω +/- 1% non-inductive
  • Duty Cycle: 1 pulse every 20s

Defib-proof and Energy Reduction Tester, Type A Type B Outputs

The MegaPulse tester provides impulses up to 4.46kV peak 1.7kΩ load. The tester has positive and negative waveform delivery via a front panel button. Tester impedance is 39ohms. The MegaPulse tester has voltage waveform 500µs rise time, 3.4ms duration and up to 4.46kV on a 1.7kΩ load.

The MegaPulse impulse tester generates the impulse waveform only; it does not determine Passing or Failing results. It is the operator’s responsibility to monitor the output waveform and determine Passing or Failing results. In monitoring the impulse waveform, consider the following points: The Impulse waveform is high voltage and high frequency (short duration). Always ensure that the measuring instrument (usually an oscilloscope with a high-voltage probe) is rated for the voltage involved, and that the frequency response of the instrument and probe are capable of measuring the output waveform of the MegaPulse Impulse Tester. A measuring instrument or probe with a low frequency response will result in erroneous readings that could be mis-read.

Determination of Passing and Failing results can prove difficult. To obtain the most accurate results, it is generally necessary to perform multiple impulse tests on a few different test samples (that have adequate insulation to pass the impulse test). Take note of the impulse waveshape, amplitude, and duration. Also note how much variance there is in the waveshape from test to test. Also (if possible), perform impulse testing on some test samples that are known to have inadequate (or damaged) insulation. Take note of the impulse waveshape, amplitude, and duration, when an insulation breakdown occurs.

MegaPulse Defib-5PF-004

  • Output Voltage: 0 – 5000 V tolerance +/-1%
  • Main Capacitance: 0.13 uF +/-10% Type A, 0.50 uF +/-10% Type B (600,000 cycles life)
  • Voltage Control: Digital Set point adjusted by frontal VOLTAGE knob or by PC TestMinder
  • Voltage Display: 4 Digit LED Display
  • Voltage Meter resolution: 2V
  • Duty Cycle: 1 pulse every 20s
  • Selectable Input Voltage: 120V – 60Hz; 240V – 50Hz

​​

Front Panel:

1. Voltage adjust knob – adjust. set point of tester

2. Polarity switch

3. NOR REV indicator – state of polarity switch

4. Charge switch – starts the charge process

5. Charge indicator

6. Trigger switch – triggers output impulse waveform

7. Trigger indicator

8. Voltage meter – displays output voltage set point

9. 500:1 BNC – 500:1 reference

10-19. (10) Outputs – (4) Type A and (6) Type B

20. Type A selection switch

21. Emergency stop 

Rear Panel:

1. Input switch

2. Fuse B

3. Appliance inlet

4. Fuse replacement warning

5. Interlock switch – Emergency stop

6. RS-232 Interface (optional)

7. Fans 

Product General Attributes

Unique FeaturesIncreased accuracy in accordance with IEC 60601-2-25 and -2-27:2011 requirements
Product Weight50 LBS
Product Height16 IN
Product Length17 IN
Product Width17 IN

Reviews

There are no reviews yet.

Be the first to review “Compliance MegaPulse Defib-5PF4 – Defib-proof and Energy Reduction Tester, Type A Type B Outputs”

Manuals/Guides

Weight50 lbs
Dimensions17 × 17 × 16 in