Description
Hioki 3504-60 C Hi-Tester (120Hz and 1kHz) with GP-IB, Binning, Contact

- Comparator function
- Trigger synchronous output
- Store 99 sets of measurement configurations
- Measurement frequencies are JIS C 5101-1 compliant
- Perform constant voltage 2ms high capacitance MLCC measurements
- High accuracy low capacitance measurements
- 4 terminal contact checking
C HiTESTERs 3506, 3505, 3504 and 3504-10 measure the C value (electrostatic capacitance) of most types of non-electrolytic capacitors. Models 3506 and 3505 provide high accuracy even when measuring small-value capacitance, and are ideal for small-capacitance internationally-recognized standard measurements requiring 1 kHz, 100 kHz and 1 MHz measurement frequencies. Model 3504 can measure large-capacitance MLCC (multi-layer ceramic capacitors) at their rated voltage, so it is most suitable for 120 Hz and 1 kHz measurements of large-value, non-electrolytic capacitors. All of these C HiTESTERs provide measurements as fast as 2 ms, and comparator and Bin functions for easy integration into automated lines. As an economically priced dedicated tester for integration in automatic systems, Model 3504-10 is the same as the 3504 without the BIN function and GP-IB interface.
Enhanced repeat-measurement accuracy
The lowest capacitance range of Models 3506 and 3505 is 220 fF* (at 1 MHz), with greatly improved repeatability accuracy for very low-capacitance measurements. * 1 fF (femtofarad) = 10-15F
Enhanced absolute accuracy
A self-calibration function minimizes variations in measurement values due to changes in ambient temperature. In addition, the cable-length compensation function minimizes measurement errors when the measurement cable is extended from 1 to 2 meters (using 1.5D-2V cable)

Enhanced contact-checking function
Contact errors while measuring can be detected by the Chatter Detection function, Low-C Reject function, Current Detection Circuit Monitoring function and the Applied Voltage Value Monitoring function. Yield rates are improved by judging measurement-object contact errors as ERR instead of FAIL.


LCR Meters / Impedance Measurement Products /LCR Meters Template | |
|---|---|
| Type | Benchtop |
| Max Test Frequency | 1 kHz |
| Max Capacitance | 20 mF |
| Measurements Parameters | Capacitance (C), Dissipation Factor (D) |
| Auto Ranging | Yes |
| Graph Mode | No |
| Data Hold | Yes |
| In Circuit Testing | No |
Product General Attributes | |
| Warranty | 1 YEARS |
| Safety Approval | CE |
| Interfaces I/O | GPIB, RS-232 |
| Product Weight | 134 OUNCES (8.38 LBS) |
| Product Height | 3.94 IN |
| Product Length | 10.24 IN |
| Product Width | 8.66 IN |
| Data Logging | No |
| Calibration Included | None |
| Power Supply Voltage | 120V/220V Universal |
| Country of Origin | Japan |